LogTag TRIL-16U

LogTag Ultra‑Low Temperature Logger with USB (TRIL‑16U)

The LogTag TRIL-16U is a reusable ultra-low temperature logger for dry ice shipments and specialist cold chain monitoring. Its built-in USB connection generates PDF and CSV reports without additional software or a reader. Records more than 16,000 temperature readings from -85 °C to +40 °C.
Temperature Range: -85 °C to +40 °C

$247.50 incl. GST

THIS PRODUCT IS IN STOCK

Additional options

Quantity
Total Price: $247.50

Special Instructions for Maximising Battery Life

Please read these instructions before configuring the logger.

Down to -85 °C

Monitor specialist cold chains and dry ice shipments.

Direct USB downloads

Automatic PDF and CSV reports without a separate reader.

16,000+ readings

Monitor for up to 90 days at a 10-minute interval.

Ultra-low temperature monitoring without the complications

The TRIL-16U is a reusable temperature logger built for demanding environments including dry ice transport, pharmaceutical supply chains and life sciences. Its measurement range extends from -85 °C to +40 °C, helping you document the conditions experienced by sensitive products.

No separate reader required

At the end of a journey, connect the built-in USB port to a computer and retrieve a PDF or CSV report automatically. For additional configuration and analysis, use the free LogTag Analyzer software.

See the status at a glance

Flashing OK and ALERT lights give a simple visual indication of recording and alarm status. Set appropriate temperature alarm thresholds for the product being monitored.

Designed for repeated journeys

With more than 16,000 temperature readings of memory, the TRIL-16U can record approximately 90 days at a 10-minute sampling interval. Its compact, IP65-rated polycarbonate case is suited to shipping cartons and challenging cold chain applications. The multi-use battery is rated for up to 1,000 cumulative hours at -80 °C.

Who uses the TRIL-16U?

  • Dry ice shipping and ultra-low temperature transport
  • Pharmaceutical and life sciences cold chains
  • Vaccine processing and temperature-sensitive biological materials
  • Laboratories and specialist storage operations
dry ice

Key Features

  • Reusable temperature data logger with integrated USB
  • PDF and CSV report generation without additional software
  • Configurable temperature alarms and flashing status indicators
  • More than 16,000 readings of memory
  • Configurable sampling and start options
  • Compatible with free LogTag Analyzer software
  • IP65-rated enclosure

Key features

Maximising the life of your TRIL-8

As the temperature drops, so does the life of your battery.

The LogTag TRIL-8 Dry Ice Temperature Recorder is one of the few loggers that will even function at -78.5°C, the surface temperature of dry ice, but any electronic device operating at that temperature pays a price in battery life. Because of this, there are some limitations on the life and warranty of the battery.

Battery Life

The device is rated for just 1,000 hours at -80°C. You will need to carefully manage how you use the battery.

Battery Warranty

The logger’s battery is warranted for 12 months from the date of manufacture, not from the date you receive it. Bearing in mind that your logger might spend months on a warehouse shelf before you buy and begin using it, your actual warranty could be much shorter than you expect.  Any warranty claims must be accompanied by LTA data files for all trips.

Our Tips

Hibernate when not in use
This logger is designed to record the temperature of goods in transit. Don’t start the logger until you are ready to send, and hibernate it as soon as you have downloaded the journey data from it.

Configure the logger to Start from Hibernate

In the Configuration settings, under Logging Parameters, choose the option Push Button Start From Hibernation.

Allow to warm before downloading results

The TRIL-8 loggers must be allowed to warm up to room temperature before downloading the results via the LogTag interface.

Specifications

ModelTRIL-16U
Measurement range-85 °C to +40 °C
Temperature accuracyBetter than ±1 °C (-30 °C to +20 °C); ±1.2 °C (-45 °C to -30 °C and +20 °C to +40 °C); ±1.7 °C (-80 °C to -45 °C)
Recording capacity16,000+ readings; up to 90 days at 10-minute intervals
Logging intervalFrom 1 minute to several hours
Start optionsPush button or programmed date and time
IndicatorsFlashing OK and ALERT lights
ProtectionIP65
BatteryFixed 3.6 V low-temperature lithium battery
Low-temperature useUp to 1,000 cumulative hours at -80 °C
Dimensions86 × 54.5 × 8.6 mm
Weight33 g
CasePolycarbonate
[remotepage page="shipping-insurance.html"]
Latency explained

What is latency?

Latency, when it comes to temperature logging, refers to the delay between a change in temperature and when that change is detected and recorded by the logger. This delay can be due to the sensor’s design, the materials it’s placed in, or the logging interval. While it might sound like a disadvantage, latency can actually serve a useful purpose in certain environments—particularly where short-term temperature spikes are common but not harmful.

For example, in a busy commercial fridge, the temperature may briefly rise every time the door is opened. A logger with high latency won’t immediately react to these short fluctuations, helping to avoid unnecessary alarms or false data indicating a problem when there isn’t one. In these cases, a slight delay in response acts like a filter, focusing attention on real issues—such as prolonged exposure to unsafe temperatures—while ignoring the everyday ups and downs that don’t affect product quality or safety.

Differrnt connections

Connecting your logger

Depending on the type of logger you have, connecting your logger to your PC means either:

  • inserting into the Reader,
  • plugging it into a USB port, or
  • connecting it via a USB cable.
comparing gen1 and gen 2

What's so good about USB?

First Generation LogTags use a separate Reader (sometimes called a dock) for configuration and for downloading recorded data. 

The Second Generation loggers only require a USB port or cable. Apart from saving the cost of the Reader, this makes them much better suited for shipping, because the recipient doesn’t need a LogTag Reader to download a report about the shipment’s journey.

comparing gen1 and gen 2